Fig.: 3 - 46
Fig.: 3 - 47
EN
Infusomat® Space 6.0
Disassembly / Assembly
6. During installation of the complete housing front panel on
the bottom inner frame the positioning hemispheres (arrow,
Fig.: 3 -
46) of the inner frame must be adjusted in the slots
of the bottom inner frame.
7. The housing front panel is pushed on to the bottom inner
frame until stop and aligned in parallel (arrow,
Snap-in Hook
1. The snap-in hook must be inserted before mounting the
housing bottom part.
Housing Bottom Part
1. Check the correct position of the seal relative to the safety
clamp before mounting the housing bottom part.
2. Make sure that the hinge spring with the housing bottom
part is pressed in the spring holder of the housing front panel.
Note
Please pay attention to the various screw lengths when you
mount the bottom part.
3
Fig.: 3 -
47).
3 - 45