Braun Infusomat Space Service Manual page 70

Infusion pump
Hide thumbs Also See for Infusomat Space:
Table of Contents

Advertisement

3
Disassembly / Assembly
1
2
6
Fig.: 3 - 23
Legend of fig. 3 - 23:
Item Designation
1
Housing front panel, cpl.
2
Fillister head screw M3x6 TORX
3
Inner frame, cpl.
4
PCB of direction of rotation sensor
5
Screw DELTA PT 25x7 WN 5451 TORX plus 8IP
6
Bottom inner frame
3 - 24
4. Loosen four screws
frame together with the housing front panel from the bottom
inner
frame(Fig.: 3 - 23 / Item
5. Unscrew one screw
of the direction of rotation sensor for the pump drive from
the pump frame.
2
Note
The PCB of the direction of rotation sensor is connected to the
3
processor PCB via the ribbon cable.
6. Pull the lock bolt drive connector
4
processor PCB.
7. Pull the pressure sensor connector, upstream
Item
7) off the processor PCB.
8. Pull the pressure sensor connector, downstream
/ Item
3) off the processor PCB.
5
Disassembly
1. Carefully remove the inner frame
the housing front
Note
The membrane in the housing front panel must not be used again.
Always use a new membrane for assembly.
(Fig.: 3 - 23 / Item
2)and lift the inner
6).
(Fig.: 3 - 23 / Item
5) and remove the PCB
(Fig.: 3 - 22 / Item
(Fig.: 3 - 23 / Item
(Fig.: 3 - 23 / Item
1) panel.
Infusomat® Space 6.0
2) off the
(Fig.: 3 - 22 /
(Fig.: 3 - 22
3) from
EN