Visual Inspection
(Part 2 of 3)
EN
Infusomat® Space 6.0
Electrical Safety
According to IEC 60601-1 or IEC 62353
Disassembly / Assembly
Functional Inspection
Pressure switch-off, electronic:
(see TSC for limit values)
Delivery rate: 250 ml/h
Testing without drop sensor
Infusomat® Space Line, standard type
❒ Upstream
❒ Alarm
❒ Downstream
❒ Pressure stage 2
❒ Pressure stage 5
❒ Pressure stage 9
Pressure limitation, mechanical
(see TSC for limit values)
❒ Pmax
❒ Pmin
Safety clamp
(see TSC for limit values)
❒ Pmin
Flow accuracy
(see TSC for limit values)
Room temperature 20 ... 25°C, delivery rate: 250 ml/h,
measuring quantity: 25 ml
❒ Deviation within tolerance
Drop sensor (optional)
Delivery rate: 250 ml/h
❒ Drop sensor alarm passage
❒ Drop sensor alarm closure