Braun Infusomat Space Service Manual page 36

Infusion pump
Hide thumbs Also See for Infusomat Space:
Table of Contents

Advertisement

2
Unit Diagnosis / Calibration
Error Code
54
2230
Drive totphase is defective
55
2231
Temperature sensor defective
56
2236 ... 2255 Internal error
57
2260 ... 2389 TCI error
58
2290 ... 2301 Internal error
59
2320 ... 2328 RTM error
60
2340 ... 2344 INTBOL error
61
2360 ... 2363 PGM error
62
2380
Internal error
63
2400 ... 2401 TOM error
64
2420
Internal error
65
8000 ... 8049 TCI error
66
8100 ... 8102 TOM error
67
9000 ... 9999 Internal error
Table 2 - 3 Error Codes of the Control Microprocessor (Part 3 of 3)
2 - 6
Definition
Software defective or obsolete
Software defective or obsolete
Error in relation to TCI therapy
together with special settings
Error in relation to Ramp and
Tapper therapy together with
special settings
Error in relation to Interval Bolus
therapy together with special
settings
Error in relation to Program
Mode therapy together with spe-
cial settings
Error in relation to Take over
Mode together with special set-
tings
Error in relation to TCI therapy
together with special settings
Error in relation to Take over
Mode together with special set-
tings
Possible Cause
Update software, change ISP proces-
sor PCB if necessary
Update software, change ISP proces-
sor PCB if necessary
Restart the device, change settings
slightly if necessary. If the error per-
sists: Please load the history file via
HiBaSeD and inform B. Braun
Restart the device, change settings
slightly if necessary. If the error per-
sists: please load the history file via
HiBaSeD and inform B. Braun
Restart the device, change settings
slightly if necessary. If the error per-
sists: Please load the history file via
HiBaSeD and inform B. Braun
Restart the device, change settings
slightly if necessary. If the error per-
sists: Please load the history file via
HiBaSeD and inform B. Braun
Restart the device, change settings
slightly if necessary. If the error per-
sists: Please load the history file via
HiBaSeD and inform B. Braun
Restart the device, change settings
slightly if necessary. If the error per-
sists: Please load the history file via
HiBaSeD and inform B. Braun
Restart the device, change settings
slightly if necessary. If the error per-
sists: Please load the history file via
HiBaSeD and inform B. Braun
Fault Clearance
Infusomat® Space 6.0
EN