3
Disassembly / Assembly
1
Fig.: 3 - 19
Legend of fig. 3 - 19:
Item Designation
1
Snap-on openings
3 - 20
3. Loosen the snap-on mechanism of the housing upper part
(Fig.: 3 - 19 / Item
1) carefully. To do so, a screwdriver can be
inserted in the two openings between the housing front panel
and the housing top.
Note
It is also possible to remove the housing upper part by gripping
into the battery compartment and tilting the upper part forward.
4. Remove the housing upper part vertically to the top.
Note
Pay attention to the length of the connection cable to the battery
compartment and to the connectors P2 and P3 when dismounting
the housing. Do not pull the connectors together with the upper
part out of the bottom part. Pay special attention to the loud-
speaker and the length of the connection cable to the battery
compartment.
Infusomat® Space 6.0