Braun Infusomat Space Service Manual page 119

Infusion pump
Hide thumbs Also See for Infusomat Space:
Table of Contents

Advertisement

2
Fig.: 6 - 6
2
Fig.: 6 - 7
EN
Infusomat® Space 6.0
Procedural Instructions on the TSC
4. Vent test setup, position of the three-way cocks please see
Fig.: 6 -
system (without pressure gauge).
Upstream
5. Push roller clamp of the Infusomat® Space Line up to approx.
30 cm before the pump and close the roller clamp or pinch
1
the Infusomat® Space Line between the container and 30 cm
before the pump. An alarm is output.
6. Open roller clamp.
Downstream
7. Switch over three-way valve, please see
er towards the pressure gauge.
Read off maximum value on the pressure gauge upon an
alarm and before an automatic pressure reduction and com-
pare with the specifications in the TSC.
8. Check all pressure stages listed in the TSC and document val-
1
ues.
Pressure Limitation, Mechanical
The optional drop sensor must not be connected.
Note
Do not use an Infusomat® Space Line with an upstream "Y" site.
The back check-valve associated with the upstream "Y" site may
skew Mechanical Pressure Limitation test results.
P
and P
max
1. Plug service connector Space on connector "P2" so that the
electronic pressure cut-off is deactivated.
2. Input a delivery rate according to the TSC, start infusion and
deliver first of all in to an open system (without pressure
gauge), position of the three-way cocks please see
3. Switch over three-way valve, please see
er towards the pressure gauge.
Read off the corresponding value on the pressure gauge and
compare with the specifications of the TSC for P
6, start infusion and deliver first of all in to an open
min
6
Fig.: 6 -
7, and deliv-
Fig.: 6 -
6.
Fig.: 6 -
7, and deliv-
und P
.
max
min
6 - 7