Temp Alarm Message - EDAN M3 User Manual

Vital signs monitor
Hide thumbs Also See for M3:
Table of Contents

Advertisement

M3 Vital Signs Monitor User Manual
The alarm limits are as follows:
Patient Type
ADU
Oral/Axillary/Rectal
PED
Oral/Axillary/Rectal

10.1.4 TEMP Alarm Message

Tables below describe the possible physiological alarms and technical alarms occurring during
TEMP measurement.
Physiological alarms:
Message
TEMP HIGH
TEMP LOW
Technical alarms:
Message
Temp
COMM
STOP
Temp
exceed
limit
No
TEMP
SENSOR
Ambient temp too
high
Ambient temp too
low
Measure position
Measuring value of TEMP is above upper alarm limit.
Measuring value of TEMP is below lower alarm limit.
Cause
TEMP
module
failure
communication
failure.
The TEMP value is
beyond the range of
+25℃ ~ +45 .
TEMP sensor is not
connected
to
TEMP module.
The
Sensor
temperature is higher
than +40℃
The
Sensor
temperature is lower
than +10℃
ALM HI
+42 ℃ (+107.6 ℉)
+42 ℃ (+107.6 ℉)
Cause
Alarm Level
or
High
Med
the
Low
Low
Low
- 82 -
TEMP Monitoring (Optional)
ALM LO
+35.5 ℃ (+95.9 ℉) +0.1 ℃
+35.5 ℃ (+95.9 ℉) +0.1 ℃
Alarm Level
What to do
Stop using measuring function
of
TEMP
module;
biomedical
engineer
Manufacturer's service staff.
Put the sensor into the sensor
bracket,
take
it
measure again.
Connect the sensor and the
monitor well, and measure
again.
Put the sensor into the sensor
bracket, measure again after
the
ambient
temperature
reaches normal value.
Step
Med
Med
notify
or
out
and