Reprocessing Parts Mechanically - Weinmann MEDUMAT Standard2 Instructions For Use Manual

Ventilator
Hide thumbs Also See for MEDUMAT Standard2:
Table of Contents

Advertisement

4. Draw the immersion disinfection solution up through the tube
5. Disconnect the disposable syringe from the tube.
6. Immerse the tube in the disinfection solution.
7. After the exposure time: Rinse the tube with drinking-water
8. Repeat the process based on this principle for each tube.
9. Allow the tubes to dry out completely.
10. If necessary: Dry tubes with sterile compressed air or medical
11. Check the tubes for residue and any remaining dirt.
12. If visible dirt remains: Repeat immersion disinfection.
Result
The reusable measuring circuit has been immersion disinfected.

8.8 Reprocessing parts mechanically

8.8.1 Reprocessing parts in a washer and disinfector
Alternatively to manual cleaning and disinfection, certain parts can
also be machine cleaned and disinfected.
Requirement
The parts have been prepared for mechanical reprocessing
(see "8.3 Preparing hygienic reprocessing", page
1. The parts approved for machine cleaning and disinfection can
into the disposable syringe by means of suction until both are
completely full.
When doing so, note:
All surfaces and lumens must be completely wetted.
Ensure compliance with the exposure time specified in the
cleaning and disinfection plan.
quality at least eight times with a syringe.
When doing so, note: Only rinse in one direction.
oxygen.
be found in the cleaning and disinfection plan
(see "8.11 Cleaning and disinfection plan", page
8 Hygienic reprocessing
MEDUMAT Standard
175).
196).
2
EN
187