9 Function check
Display
Device is ready for use
Device is ready for use
Service due in %1 days
Device is ready for use
Service due
Maintenance symbol is
flashing in the display
(only in the start menu)
Device is ready for use
Check or replace the
FlowCheck sensor
Maintenance symbol is
flashing in the display
(only in the start menu)
Device is not ready for use
208
EN
MEDUMAT Standard
14. Proceed with the device according to the following table:
Meaning
Function check passed.
Function check passed
≤ 60 days until expiry of
the maintenance interval
Function check passed
Maintenance interval
expired
Function check passed
Total useful life of the
FlowCheck sensor has
been exceeded
Function check failed.
The message Check or replace the FlowCheck sensor can also
appear with a notice that maintenance is required.
15. Press the navigation knob finish.
16. Switch off the device.
17. Close the oxygen cylinder.
2
Action
Use device without restriction.
Device can be used
without restrictions
until expiry of the
maintenance interval.
To continue using the
device without
restrictions: Contact
WEINMANN Emergency
or a technician
authorized by
WEINMANN Emergency
in good time for
maintenance.
To continue using the device
without restrictions: Contact
WEINMANN Emergency or
a technician authorized by
WEINMANN Emergency for
maintenance.
To continue using the device
without restrictions: Check or
replace the FlowCheck sensor.
Take action
(see "9.4 Failed
function check", page
209).