Weinmann MEDUMAT Standard2 Instructions For Use Manual page 49

Ventilator
Hide thumbs Also See for MEDUMAT Standard2:
Table of Contents

Advertisement

Status indicator
Device
If the device is switched on, you can see the battery status on
the display:
Status indicator
4 Preparation and operation
Explanation
1 LED is flashing
No LEDs are lit
Red fault indicator
is lit
Meaning
Battery status > 90 %
Battery status approx. 60 % - 90 %
Battery status approx. 40 % - 60 %
Battery status approx. 10 % - 40 %
Battery status < 10 %
The last remaining segment in
the battery status symbol is red.
The message Battery weak
appears in the display.
MEDUMAT Standard
Meaning
Battery status < 10 %
Battery is deeply discharged.
Charge battery in the device
for 24 hours. After 24 hours:
Green LED is lit: Battery
fully charged and ready
for use.
Red LED or no LED is lit:
Battery defective.
Replace battery.
Battery defective.
Replace battery.
2