4 Preparation and operation
Requirement
Result
4.7.12 Changing the ventilation mode in CPR mode
Requirement
84
EN
MEDUMAT Standard
If the trigger is still not functioning properly in CPR CCSV mode
in spite of reducing the trigger and increasing the PEEP, you can
switch from CPR CCSV mode to CPR IPPV mode
(see "4.7.12 Changing the ventilation mode in CPR mode",
page
84).
Pausing ventilation in CPR CCSV mode
During the analysis of the defibrillator, you can pause ventilation
in order to avoid ventilation artifacts in the analysis.
The device is switched on
The CPR CCSV mode is set.
1. Pause chest compressions.
The ventilation is paused for as long as the chest compressions
are continued (up to the set back-up ventilation time at most).
This chapter describes how you can switch between CPR Manual,
CPR IPPV and CPR CCSV in CPR mode.
The device is switched on.
A CPR mode is set.