7 Hygienic reprocessing

7.8 Reprocessing parts mechanically

Permitted parts
Requirements
88
EN
MEDUVENT Standard
As an alternative to manual cleaning and disinfection, certain parts
can also be cleaned and disinfected by mechanical means.
Part
Hook and loop strap with
clip
Reusable breathing circuit
Ventilation hose
Patient valve
Elbow
Protective cap
Protective sleeve
Reusable measuring circuit,
comprising:
PEEP control tube
Pressure measuring
tube
Flexible oxygen tube
Measuring circuit
connector
The parts have been prepared for mechanical reprocessing
"7.2 Preparing for hygienic reprocessing", page
1. For parts approved for mechanical cleaning and disinfection,
refer to the cleaning and disinfection plan
and disinfection plan", page
Mechanical reprocessing
Wash with Derval SOLO and Ottalin PERACET at
up to 70 °C
Cleaning:
®
neodisher
MediClean forte (Dr. Weigert):
0.5 %, 55 °C, 10 minutes
Thermal disinfection: 90 °C, 5 min
(corresponds to A0 value 3000)
Wash at 60 °C in an industrial washing
machine
Cleaning agent: Derval SOLO (RKI) (Kreussler)
Dose: 2 ml/l
and
disinfectant:
Ottalin PERACET (Kreussler)
Dose: 2 ml/l
Exposure time: 10 min, type AB
Cleaning agent:
®
neodisher
MediClean forte (Dr. Weigert):
0.5 %, 55 °C, 10 minutes
Thermal disinfection: 90 °C, 5 min
(corresponds to A0 value 3000)
92).
(see
74).
(see "7.11 Cleaning