Specification
Minimum limiting pressure (pLim, min) 10 mbar. The device does not generate active negative pressure.
Maximum limiting pressure
(pLim, max.)
Means for limiting pressure
Means for ensuring the minimum value Pressure control
PEEP
PEEP CCSV (only with CCSV option)
Pressure support ΔpASB (only with
Flow measurement + ASB option)
Pressure ramp (only with Flow
measurement + ASB option)
Trigger (permanently set)
Trigger can be set (only with Flow
(1)
measurement + ASB
Trigger can be set
(only with CCSV option)
Device
65 mbar
Pressure control
0 mbar to 30 mbar (±3 mbar or ±15 %)
0 mbar to 5 mbar (±3 mbar or ±15 %)
0 mbar to 35 mbar (±3 mbar or ±15 %) above PEEP
Steep
Averaging
Flat
Inspiratory trigger:
-1.2 mbar at PEEP > 0
-0.8 mbar at PEEP = 0
(1)
Expiratory trigger:
30 % of the maximal flow
Triggering via internal sensors
Triggering via FlowCheck sensor
Inspiratory trigger (3 levels):
Level 1: Sensitive, corresponds to approx. 3 l/min
Level 2: Medium sensitivity, corresponds to approx. 5 l/min
Level 3: Insensitive, corresponds to approx. 10 l/min
Expiratory trigger (3 levels):
Level 1: Long ASB breath, corresponds to approx.
10 % of the inspiratory maximum flow
option)
Level 2: Medium ASB breath, corresponds to approx.
35 % of the inspiratory maximum flow
Level 3: Short ASB breath, corresponds to approx.
70 % of the inspiratory maximum flow
Inspiratory trigger (units):
1 l/min to 15 l/min
Expiratory trigger (units)
5 % flow max. to 80 % flow max.
Inspiratory trigger (5 levels):
Level 1: Very sensitive (only slight pressure on the chest
is necessary to trigger a mechanical breath)
to
Level 5: Insensitive (firm pressure on the chest is necessary
to trigger a mechanical breath)
MEDUMAT Standard
14 Appendix
2
EN
235