Weinmann MEDUMAT Standard2 Instructions For Use Manual page 15

Ventilator
Hide thumbs Also See for MEDUMAT Standard2:
Table of Contents

Advertisement

2.3.2 Power supply
Warning
Risk of injury due to missing, flat or defective battery!
A missing, flat or defective battery prevents treatment.
 Only operate the device with a charged battery.
 Keep an alternative ventilation unit at the ready.
Treatment prevented by defective power cord or power
supply unit!
A defective power cord or power supply unit prevents the battery
in the device from charging and thus impairs the operational
readiness of the device.
 Inspect the power cord and power supply unit regularly.
 Only operate the device with a charged battery.
 Keep an alternative ventilation unit at the ready.
Risk of injury due to electric shock when connecting an
incorrect power supply unit to the line power!
The power supply unit contains a safety device to prevent electric
shock. The use of a non-original power supply unit may result in
injury to the user.
 Only operate the device on line power using the power supply
unit recommended by WEINMANN Emergency.
Risk of injury due to incorrect maintenance of the Li-ion
battery!
An open, non-original or damaged Li-ion battery can lead to
excessive temperatures, fire or explosions. This can lead to injuries
to the user, patient or third parties.
 Only operate the device using the battery recommended by
WEINMANN Emergency.
 Only use an unopened, undamaged battery.
Caution
Risk of injury from touching the contacts in the battery
compartment and the patient at the same time!
The contacts in the battery compartment are live. Touching the
contacts and the patient at the same time can injure the user or
the patient.
 Do not touch the contacts in the battery compartment and the
patient at the same time.
MEDUMAT Standard
2 Safety
2
EN
15