Weinmann MEDUMAT Standard2 Instructions For Use Manual page 186

Ventilator
Hide thumbs Also See for MEDUMAT Standard2:
Table of Contents

Advertisement

8 Hygienic reprocessing
Result
8.7.1 Immerse disinfecting the reusable measuring
Requirement
186
EN
MEDUMAT Standard
5. After the specified exposure time, rinse the parts for at least
10 s and textiles for at least 2 min with drinking-water quality
in order to remove all of the disinfectant residue.
6. Allow the parts to dry fully.
7. Check the parts for residue and any remaining dirt.
8. If dirt is visible: Repeat the cleaning and disinfection.
The parts are immersion disinfected.
circuit
Risk of infection and contamination due to insufficient
hygienic reprocessing of the measuring circuit!
Rinsing the measuring circuit in alternating directions does not
guarantee any bacteria reduction and may injure the patient.
 Only rinse measuring tubes in one direction.
The principle described only applies to the following parts of the
reusable measuring circuit:
Pressure measuring tube
PEEP control tube
Measuring circuit connector
The reusable measuring circuit is disconnected from the
reusable breathing circuit
reusable breathing circuit", page
The reusable measuring circuit has been cleaned manually.
1. The disinfectants, doses and exposure time for the individual
parts can be found in the cleaning and disinfection plan
(see "8.11 Cleaning and disinfection plan", page
2. Prepare the immersion disinfection solution in accordance with
the specifications of the disinfectant agent manufacturer.
3. Connect a sterile disposable syringe (20 ml) to one end of a
tube.
2
(see "8.4 Disassembly of the
176).
196).