Risk of injury in the event of device failure resulting from
blocked suction inlets on the hygiene filter!
Blocked suction inlets on the hygiene filter may cause injury to the
patient in the event of device failure as a result of excessively high
pressures, and may prevent the patient from breathing on his/her
own.
 Always keep the suction inlets on the hygiene filter clear.
Risk of injury due to sparks during defibrillation in the
presence of oxygen and combustible materials!
In the event that a ventilator and defibrillator are used at the same
time, defibrillation in an oxygen-enriched atmosphere and in the
presence of combustible materials (e.g., textiles) combined with
sparks generated by the defibrillation may cause explosions and
fire, which may result in injury to the patient, user or bystanders.
 During defibrillation, only use adhesive electrodes or ensure
that the oxygen-air mixture coming from the patient valve
flows away from the torso of the patient.
Risk of injury due to concealed alarm!
A concealed alarm light, loudspeaker and display will prevent the
user from noticing any alarms and reacting to dangerous
situations. This may result in injury to the patient.
 Always keep the alarm (alarm light, loudspeaker and display)
free.
 Do not operate the device in a closed bag.
Risk of injury if an incorrect volume is applied in hyperbaric
environments!
Use of the device in hyperbaric environments (pressure chambers)
leads to the application of incorrect volumes and may result in an
injury to the patient.
 Do not use the device in hyperbaric environments.
Risk of injury from operating the device, accessories and
components outside of the prescribed ambient conditions!
Use of the device, accessories and components outside of the
prescribed ambient conditions may mean that tolerances are not
adhered to and result in device failure and injury to the patient.
 Only operate the device and components within the prescribed
ambient conditions