4 Preparation and operation
Requirement
82
EN
MEDUMAT Standard
Treatment delays due to triggered alarms for spontaneously
breathing patients which do not apply in the this application
scenario!
In CPR CCSV mode, triggered etCO
user and result in delays in treatment.
 Switch off etCO
operator menu.
Treatment delay due to impaired detection of chest
compressions due to obstructed airways!
In CPR CCSV mode, chest compressions are detected as a result of
pressure changes in the airways. If the airways are obstructed,
this detection is impaired, which can lead to delays in treatment.
 Check the setting of the PEEP and the trigger in CPR CCSV
mode and adjust if required.
The device is switched on.
The CCSV option is activated.
1. Briefly press the CPR button
Depending on the preset in the operator menu, the device
switches to either the CPR CCSV or the CPR Manual mode
(Operator menu | Presets patient | CPR mode | Start mode).
2. If
is already set in the operator menu: Continue
ventilation via chest compressions in CPR CCSV mode:
or
If Manual is set in the operator menu: Switch to
with the right-hand navigation knob
the ventilation mode in CPR mode", page
2
alarms can confuse the
2
alarms for CPR CCSV mode in the
2
.
CCSV
(see "4.7.12 Changing
84).