6. Immerse the tube in the solution for disinfection.
7. Once the exposure time has elapsed: Rinse the tube at least
8. Following this principle and carry out the procedure for each
9. Allow the tubes to dry completely.
10. If necessary: Dry the tubes with medical compressed air or
11. Check the tubes for residues and residual soiling.
12. If visible soiling is present: Repeat disinfection by immersion.
Result
The reusable measuring circuit has been disinfected by immersion.

8.8 Reprocessing parts mechanically

As an alternative to manual cleaning and disinfection, certain parts
can also be cleaned and disinfected by mechanical means.
Permitted parts
When doing so:
All surfaces and lumina must be wetted completely.
Observe the exposure time specified in the cleaning and
disinfection plan.
8 times in water of potable quality using the syringe.
When doing so: Flush only in one direction.
tube.
medical oxygen.
Part
Hook and loop strap
with clip
Reusable breathing circuit
Ventilation hose
Patient valve
Elbow
Protective cap
8 Hygienic reprocessing
Mechanical reprocessing
Wash with Derval SOLO and Ottalin PERACET
at up to 70 °C
Cleaning:
®
neodisher
MediClean forte (Dr. Weigert):
0.5 %, 55 °C, 10 minutes
Thermal disinfection: 90 °C, 5 min
(corresponds to A0 value 3000)
MEDUVENT Standard
EN
101