9.7 Rectifying leaks in the system

Requirement
1. Prepare a soapy solution using unperfumed soap.
Damage to the device caused by ingress of liquids!
Ingress of liquids may damage the device, components and
accessories.
 Do not immerse the device, components or accessories in
2. Wet all screw connections and hoses/tubes with the soapy
3. In the event of a leakage: Close the valve on the oxygen
4. Briefly press the On/Off button
5. Press and hold the On/Off button
6. Replace leaky components.
7. Check the system for leaks once more
8. If necessary: Look for other leaks and replace leaky
9. If the leak cannot be rectified, have the device repaired.
Result
The leak in the system has been rectified.
All screw connections are tightened.
All hoses/tubes are securely connected.
There is a leak in the system.
liquids.
solution.
Bubbles will form if a leak is present.
cylinder.
without an oxygen supply.
The remaining oxygen is flushed out of the device.
to switch off the device.
system for leaks", page
components.
and operate the device
for at least 2 seconds
(see "9.6 Checking the
210).
MEDUMAT Standard
9 Function check
2
EN
211