6 Operator menu
Parameter
Infant/Child/Adult
Emergency mode
IPPV
BiLevel + ASB
CPAP + ASB
PCV
aPCV
PRVC
SIMV + ASB
S-IPPV
RSI
CPR
CPR mode
Start mode
Metronome
Metronome freq.
/ (only with
etCO
2
Capnography option)
134
EN
MEDUMAT Standard
Possible values
IPPV
BiLevel + ASB
CPAP + ASB
PCV
SIMV + ASB
S-IPPV
PRVC
aPCV
Factory settings
(see 14.1.10, p. 257)
Ventilation parameters
Alarm limits
Manual
CPR IPPV
CPR CCSV
100/min - 120/min
2
Description
Here you can select an emergency
ventilation mode per patient group.
Here you can make the presets for
the ventilation mode.
Here you can set with which
submenu the CPR mode should
be started. If the CCSV option is
enabled, once the patient group
Child or Adult is selected, the device
always starts in continuous mode
with CPR CCSV.
Here you can activate or deactivate
the audio output of the metronome.
Here you can set the frequency of the
metronome tone.
Here you can determine whether an
alarm should be output in the event
of rising or dropping end-expiratory
CO
.
2