Weinmann MEDUMAT Standard2 Instructions For Use Manual page 253

Ventilator
Hide thumbs Also See for MEDUMAT Standard2:
Table of Contents

Advertisement

Guidelines and manufacturer's declaration – electromagnetic immunity
2
MEDUMAT Standard
The customer or the user of the MEDUMAT Standard
an environment.
Immunity tests
Electrostatic
discharge (ESD)
acc. to IEC 61000-4-2
Bursts acc. to
IEC 61000-4-4
Surges acc. to
IEC 61000-4-5
Voltage dips, short-
term power failures
and fluctuations in
the supply voltage
acc. to IEC 61000-4-11
Note:
is the AC voltage in the mains prior to application of the test level.
U
T
Interference pulse
on motor vehicle
supply lines acc. to
ISO 7637-2
is intended for operation in the electromagnetic environment specified below.
IEC 60601 test
level
± 8 kV contact
discharge
± 15 kV air discharge
± 2 kV for mains
power lines
± 1 kV for input and
output lines
± 1 kV voltage
Phase-to-phase
± 2 kV voltage
Phase-to-earth
0 % U
; ½ cycle at 0,
T
45, 90, 135, 180, 225,
270 and 315 degrees,
0 % U
,1 cycle and
T
70 % U
, 25/30 cycles,
T
Single phase:
at 0 degrees,
0 % U
,
T
250/300 cycles
Test pulse 1, 2a, 2b,
3a, 3b and 4
2
device must ensure that it is also used in such
Compliance level
± 8 kV contact
discharge
± 15 kV air discharge
± 2 kV for mains
power lines
± 1 kV for input and
output lines
± 1 kV voltage
Phase-to-phase
± 2 kV voltage
Phase-to-earth
0 % U
; ½ cycle at 0,
T
45, 90, 135, 180, 225,
270 and 315 degrees,
0 % U
,1 cycle and
T
70 % U
, 25/30 cycles,
T
Single phase:
at 0 degrees,
0 % U
,
T
250/300 cycles
Test pulse 1, 2a, 2b,
3a, 3b and 4
MEDUMAT Standard
14 Appendix
Electromagnetic
environment – guidelines
Floors should be made of wood or
concrete, or covered with ceramic
tiles. If the floor is covered with
a synthetic material, the relative
humidity must be at least 30 %.
The quality of the supply voltage
should correspond to that of
a typical business or hospital
environment.
The quality of the supply voltage
should correspond to that of
a typical business or hospital
environment.
The quality of the supply voltage
should correspond to that of
a typical business or hospital
environment. If the user of the
2
MEDUMAT Standard
continuing function even if the
power supply is interrupted,
we recommend running
2
MEDUMAT Standard
a fully charged battery.
The vehicle on which
2
MEDUMAT Standard
mounted must be E1-certified.
2
EN
demands
with
is to be
253