Weinmann MEDUMAT Standard2 Instructions For Use Manual page 199

Ventilator
Hide thumbs Also See for MEDUMAT Standard2:
Table of Contents

Advertisement

Part
Manual cleaning
Disposable breathing circuit
Disposable
breathing circuit
Disposable
measuring circuit
Hose clips
FlowCheck sensor
Disposable, do not reuse. Dispose of correctly
(disposable)
Water filter
CO
measuring
2
tube (only with
Capnography
option)
FlowCheck sensor
Wipe down with
connection line
neodisher
MediClean forte
FlowCheck sensor
(Dr. Weigert).
connection line
with MEDUtrigger
Dosage: 10 ml/l,
Duration: treat all
MEDUtrigger with
surfaces at least
connection line
twice until visibly
clean
Wipe
disinfection
®
Wipe down
with Incidin™
OxyWipe S (Ecolab).
Exposure time:
5 min
The applicable instructions are those in the instructions for use
from the manufacturers of the individual components or parts.
Observe these instructions for use.
8 Hygienic reprocessing
Immersion
disinfection
(see 13, p.
230).
Not permitted
MEDUMAT Standard
Mechanical
reprocessing
(Disinfection)
Not permitted
2
EN
199