8 Hygienic reprocessing

8.9 Preparing parts for reuse

Requirement
Result

8.10 Assembling reusable breathing circuit

Requirement
190
EN
MEDUMAT Standard
The parts have been subjected to hygienic reprocessing according
to the cleaning and disinfection plan.
1. Check all parts for damage due to use (e.g., tension cracks and
cable breakage).
2. Replace damaged parts.
3. Assemble the reusable breathing circuit
reusable breathing circuit", page
4. Mount the accessories.
5. Reconnect the power supply
supply", page
45).
6. Perform a function check
7. Place parts in storage in accordance with the storage
conditions
(see "14.1 Technical data", page
The parts are ready for use again.
The images in this subchapter show all the possible components
of the reusable breathing circuit. Depending on the system type,
your reusable breathing circuit may not include certain
components
(see "3.5.2 Reusable breathing circuit and disposable
breathing circuit", page
The reusable breathing circuit is disassembled.
2
(see "8.10 Assembling
190).
(see "4.2 Connecting to a power
(see "9 Function check", page
35).
200).
231).