7 Description of the modes
7.3.4 CPAP + ASB mode
Pressure
pMax
CPAP / PEEP
148
EN
MEDUMAT Standard
Description
Abbreviation
Long form
Type
Requirement
Ventilation parameters
Left-hand navigation knob
Central navigation knob
Right-hand navigation knob
CPAP without pressure support
1/Freq. (spontaneous)
The CPAP + ASB mode can be separated into its individual
elements:
The set value CPAP/PEEP is used to increase the pressure level
of respiration in order to raise the functional residual capacity
(FRC) of a spontaneously breathing patient.
2
CPAP + ASB
Continuous Positive Airway Pressure +
Assisted Spontaneous Breathing
Pressure-controlled
Flow measurement + ASB option is
activated
InTr
ΔpASB
CPAP + ASB
CPAP with pressure support ASB
1/Freq. (spontaneous)
PEEP
pMax
ExTr
Emergency mode
ΔpASB
Time