Calibrate Speed

From the Operating screen, press the down arrow, up arrow and Silence
button in that order, to access the Configuration menu.
From the screen at right, press the
down arrow twice.
Press the Speed button beneath
RPM at this screen.
The value shown on this screen is
the present Speed setpoint. Using
the up and down arrows, increase
or decrease the platform speed
until the reading on an
independent, accurate speed measuring device matches the shaker speed
setpoint.
When set, press the Speed button to save the setting. The display will
return to the Calibrate - RPM Temp screen. Or, if nothing is pressed for
about fifteen seconds, the display will return to the Operating Screen and
the setting will be automatically saved to memory.
Calibrate the
Begin by pressing the down arrow, the up arrow, the Silence button. Then
press the down arrow twice. The screen below will appear on the display:
Temperature
Press the Temperature button
beneath Temp.
Using the up and down arrows,
increase or decrease the
temperature value to match an
independent, accurate temperature
measuring device. When selected,
press the Time, Speed, or Temp button to save the setting. The display will
return to the Calibrate - RPM Temp screen. (Or, if nothing is pressed for
about fifteen seconds, the display will return to the Operating Screen and
the setting will be automatically saved to memory.)