9337 Instrument Fault 310V Psu; 9339 Mono Resetting. Please Wait; 9415 Instrument Fault: Lamp Code; 9422 Instrument Fault: Lamp Current - Agilent Technologies 55B AA User Manual

Hide thumbs Also See for 55B AA:
Table of Contents

Advertisement

Troubleshooting/Errors
Suggested action:
Check that the optical light path is not obstructed in the sample
compartment. Correct any of the possible causes listed above. Exit,
then re-enter the Optimize page/window. This will cause the
instrument to retry the peaking.

9337 Instrument Fault 310V PSU

Instrument Fault. The instrument detected a problem with the 310
volt supply on the control board. A service call may be required.

9339 Mono resetting. Please wait

This message is for information only. It can only occur when the
instrument is just switched on. If this message appears then the
instrument will take several minutes to find its wavelength reference.
The instrument will function correctly after the delay. If this message
continues to occur and the startup delay is a problem, then a service
call will be required.

9415 Instrument Fault: Lamp code

The instrument could not recognise the lamp code for one of the
installed hollow cathode lamps.
Possible causes include:

9422 Instrument Fault: Lamp current

The instrument could not detect any lamp current for the selected
hollow cathode lamp.
Possible causes include:
130
One of the installed hollow cathode lamps is faulty
A electronic fault in the instrument's lamp recognition circuity.
No hollow cathode lamp is installed in the selected lamp position
Faulty hollow cathode lamp
Faulty lamp supply circuits.
Agilent 55B AA Spectrometer User's Guide

Advertisement

Table of Contents
loading

Table of Contents