1.2.5 ACCURACY

A.
Low Flow Cutoff - 2% of Full Scale Differential Pressure of lowest
range transducers for each tube.
B.
Density - determined from the Frequency Densitometer input; ±0.1%
of densitometer span for a minimum span of 0.2 LBF3.
C.
Rate Determination - is ±0.1% of full scale.
D.
Temperature Coefficient for Totals - is 0.005%/F.
14
Up Arrow (↑) - Pressing ↑ results in the following actions by
the computer:
Reading data - ↑ causes the computer to step back to the
a.
previous data code.
corresponding to Read Code 2 is being viewed, pressing
↑ causes the computer to display the data corresponding
to Read Code 1.
Entering data - ↑ indicates to the computer that the data
b.
to follow is an exponent (e.g., 2 ↑ 5= 2 x 10
Down Arrow (↓) - Causes the computer to step to the next data
code.
Print (PRNT) - Pressing PRNT sends operator selected data
output to an external printer.
For example, if the data
SECTION 1
5
=200,000).