__________________________ MODEL 2234 DIGITAL FLOW COMPUTER
Display
1. CNFIG
2. DENTYP
3. ENTER IE
4. ENTER TFS
5. ENTER TZ
6. ENTER DFS
7. ENTER DZ
8. ENTER PFS
9. ENTER PZ
10. ENTER MFS
11. ENTER DOF
12. ENTER DOZ
13. ENTER TK
14. ENTER LF1
15. ENTER LK1
16. ENTER HF1
17. ENTER ID1
18. ENTER OD1
19. ENTER TL1
20. ENTER PA1
21. ENTER PT1
22. ENTER LA1
23. ENTER LT1
Repeat steps 14-23 as required to comply with the number of runs identified by the configuration
used in Step 1.
26. READY
The computer is ready for flow computations if further data entries for options are
not required.
136
Display Definition
Enter System Configuration Code Number (1-11)
Enter Densitometer Type (1,2,3,4 or 5)
Enter Isentropic Exponent
Enter Full scale for Measured Temperature in degrees F
Enter Zero Scale for Measured Temperature in degrees F
Enter Full Scale for Measured Density in Pounds per Cubic Foot
Enter Zero Scale for Measured Density in Pounds per Cubic Foot
Enter Full Scale for Measured Static Pressure in PSIA
Enter Zero Scale for Measured Static Pressure in PSIA
Enter Full Scale Mass for Station Rate in LB/HR
Enter Full Scale Density Output in LB/CF
Enter Zero Scale Density Output in LB/CF
Enter Station Totalizing Factor
Enter Low Flow Cutoff for Line 1 (InH20)
Enter Totalizing Factor for Line 1
Enter Full Scale Measured Differential Pressure for Line 1 in Inches of
Water
Enter Inside Diameter of Meter Tube for Line 1 in Inches
Enter Orifice Diameter of Meter Tube for Line 1 in Inches
Enter Pressure Tap Location for Line 1 (1=Upstream, 2=Downstream)
Enter Line 1 Plate Expansion Coefficient
Enter Line 1 Plate Measurement Temperature (DEGF)
Enter Line 1 Pipe Expansion Coefficient
Enter Line 1 Pipe Measurement Temperature (DEGF)
SECTION 3