General Procedures and Functional Checks
4-3-9 2D Mode (B mode) Checks
4-3-9-1 Introduction
4-3-10 2D Mode Screen Example

4-3-11 Preparations

4-3-12 Adjust the 2D mode controls

4-26
The 2D Mode is the system's default mode.
Connect one of the probes.
Turn ON the scanner.
The 2D Mode is displayed (default mode).
Press B Mode on the Operator Panel to access 2D mode.
These Image Controls are used to optimize the 2D picture.
Verify that all the listed controls are working as intended:
Use Gain and TGC controls to optimize the overall image
together with the Power control.
Use Depth to adjust the range to be imaged.
Use Focus to center the focal point(s) around the region of
interest.
Use Frequency (move to higher frequencies) or Line
Density (move to higher line density) to increase resolution
in image.
Use Frequency (move to lower frequency) to increase
penetration.
Use the Rejection control to reduce noise in the image.