General Procedures and Functional Checks
4-3-9 2D Mode (B mode) Checks
4-3-9-1 Introduction
4-3-10 2D Mode Screen Example
4-3-11 Preparations
4-3-12 Adjust the 2D mode controls
4-26
The 2D Mode is the system's default mode.
•
Connect one of the probes.
•
Turn ON the scanner.
The 2D Mode is displayed (default mode).
Press B Mode on the Operator Panel to access 2D mode.
These Image Controls are used to optimize the 2D picture.
Verify that all the listed controls are working as intended:
•
Use Gain and TGC controls to optimize the overall image
together with the Power control.
•
Use Depth to adjust the range to be imaged.
•
Use Focus to center the focal point(s) around the region of
interest.
•
Use Frequency (move to higher frequencies) or Line
Density (move to higher line density) to increase resolution
in image.
•
Use Frequency (move to lower frequency) to increase
penetration.
•
Use the Rejection control to reduce noise in the image.