4-3-22 PW Doppler Mode Checks

4-3-22-1 Introduction
4-3-22-2 Preparations
4-34
PW Doppler are used to measure velocity (most often in blood).
Doppler mode can be done with a special pencil probe or with
an ordinary probe. By using an ordinary probe, you can first
bring up a 2D picture for navigation purpose and then add PW
Doppler.
Connect one of the probes to the scanner.
See
4-3-27 'Probe/Connectors Check' on page 4-39
about connecting the probes.
For available probes, see
Turn ON the scanner
The 2D Mode window is displayed (default mode).
If needed, adjust the Display's Brightness and Contrast
setting.
Press PW to start Pulsed Wave Doppler (PW).
Use the trackball to select the Area of Interest (Sample
Volume) in PW.
LOGIQ V3/V5/V5 Expert – Basic Service Manual
9-3-2 'Probe' on page
5726264-100 English Rev.8
for info