10-7-6-6 Data Sheet for Transducer Source Leakage Current
CAUTION
Table 10-17:
Unit Power
ON
NORM
ON
NORM
ON
REV
ON
REV
OFF
NORM
OFF
NORM
OFF
REV
OFF
REV
LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
The test passes when all readings measure less than the values
shown in
Table 10-13 on page
Inspection Certificate.
Equipment damage possibility. Never switch the Polarity and
the status of Neutral when the Ultrasound system is powered
ON. Be sure to turn the Ultrasound system power OFF before
switching them using the POLARITY switch and/or the
NEUTRAL switch. Otherwise, the Ultrasound system may be
damaged.
Typical Data Sheet For Transducer Source Leakage Current
Transducer Tested:
Tester Power Polarity
Switch
Electrical safety tests
10-21. Record all data on the PM
Tester GROUND or
NUETRAL Switch
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
Measurement
10-29