General Procedures and Functional Checks
4-2-2-2 Connect AC (mains) Power to LOGIQ V3/V5/V5 Expert
4-6
Connecting AC Power to the LOGIQ V3/V5/V5 Expert
ultrasound unit, involves preliminary checks of the power cord,
voltage level and compliance with electrical safety requirements.
1. Ensure that the wall outlet is of appropriate type, and that
the Circuit Breaker is turned off.
2. Uncoil the power cable, allowing sufficient slack so that the
unit can be moved slightly.
3. Verify that the power cable is without any visible scratches
or any sign of damage.
4. Verify that the on-site mains voltage is within the limits
indicated on the rating label near the Circuit Breaker on the
rear of the unit.
5. Connect the Power Cable's female plug to the Power Inlet at
the rear of the unit.
6. Lock the plug in position with the Retaining Clamp (ACC
Clamp).
7. Verify that the Mains Power Circuit Breaker is in OFF
position, if not, switch it OFF.
8. Connect the Power Cable's other end (male plug) to a
hospital grade mains power outlet with the proper rated
voltage, and the unit is ready for Power ON/Boot Up.