Site Preparations
2-2-2-2 Environmental specifications for Ultrasound system
Operational
Temperature
10° - 40°C
50° - 104°F
Humidity
30 - 80%
non-condensing
Pressure
700 - 1060hPa
NOTE:
CAUTION
CAUTION
2-4
The system should be operated, stored, or transported within
the parameters outlined below. Either its operational
environment must be constantly maintained or the unit must be
turned off.
Table 2-2:
System Environmental Requirements
Temperature in degrees C. Conversion to Degrees F = (Degrees
C * 9/5) + 32.
Ensure that the probe face temperature does not exceed the
normal operation temperature range.
The LOGIQ V3/V5/V5 Expert system and probe connector is
not waterproof. Do not expose the device to water or any kind
of liquid.
Storage
-5° - 50°C
23° - 122°F
10 - 90%
non-condensing
700 - 1060hPa
LOGIQ V3/V5/V5 Expert – Basic Service Manual
Transport (<16hrs.)
-5° - 50°C
23° - 122°F
10 - 90%
non-condensing
700 - 1060hPa
5726264-100 English Rev.8