Replacement Procedures

8-3-4 Backing up the Patient Archive and System Configurations

CAUTION

8-3-5 Restoring up the Patient Archive and System Configurations

NOTE:
WARNING
8-6
An error, or a power loss may occur.
Always backup the Patient Archive and the Presets (System
Configurations) before loading the software!
In order to complete a successful restore of the Patient
Database, as needed after a hard disk replacement, or if all the
content on the hard disk has been erased, the images must be
moved away from LOGIQ V3/V5/V5 Expert before doing
backup of the Patient Database.
Depending on the location set-up, either move the images to a
remote server or to removable media like DVD or CD discs.
Backup the Patient Archive and System Configurations.
For instructions, please see "Data Backup and Restore" in
the Basic User Manual/User Guide.
For instructions, please see "Data Backup and Restore" in
the Basic User Manual/User Guide.
It is not suggested to manually delete the files in
D:\Idunn\target\resources\idunn\userdefs.
To avoid lacking of connecting to Local Archive, connecivity.res
and IPSave in D:\Idunn\target\resources\idunn\userdefs could
not be deleted. If they are deleted, please rewirte the serial
number.