J22 TDLAS Gas Analyzer
Product Safety
IT Security
SpectraSensors, Inc.
The J22 TDLAS Gas Analyzer is designed in accordance with good engineering
practice to meet state-of-the-art safety requirements, has been tested, and left
the factory in a condition in which it is safe to operate.
It meets general safety standards and legal requirements. It also complies with
the EU directives listed in the specific EU Declaration of Conformity. SpectraSen-
sors confirms this by affixing the CE mark to the analyzer system.
General
•
Adhere to all warning labels to prevent damage to the unit.
•
Do not operate the device outside the specified electrical, thermal and
mechanical parameters.
•
Only use the device in media to which the wetted materials have sufficient
durability.
•
Modifications to the device can affect the explosion protection and must be
carried out by staff authorized to perform such work by SpectraSensors.
•
Only open the controller cover if the following conditions are met:
- An explosive atmosphere is not present
- All device technical data is observed (see nameplate)
- An electrostatic charge (e.g., caused by friction, cleaning or maintenance)
is avoided on the attached stainless steel nameplate, if present, and on
painted metallic housings that are not integrated into the local potential
equalization (ground) system
•
In potentially explosive atmospheres:
- Do not disconnect any electrical connections while the equipment is ener-
gized.
- Do not open the connection compartment cover when energized or the
area is known to be hazardous.
•
Install the controller circuit wiring according to the Canadian Electrical Code
(CEC) respective National Electrical Code (NEC) using threaded conduit or
other wiring methods in accordance with articles 501 to 505, and/or IEC
60079-14.
•
Controller enclosures are factory sealed when using in ambient temperature
of not lower than -20 °C (operating) or -40 °C (storage).
•
Install the device according to the manufacturer's instructions and regula-
tions.
•
The flameproof joints of this equipment are outside the minimums specified
in IEC/EN 60079-1 and shall not be repaired by the user.
The provided warranty is only in effect if the analyzer system is installed and used
as described in the Operating Instructions. The analyzer system is equipped with
security mechanisms to protect it against any inadvertent changes to the system
settings. IT security measures are in line with operators' security standards and
designed to provide additional protection for the and data transfer must be imple-
mented by the operators themselves.
Protecting access via hardware write protection
Write access to the device parameters via the local display or Web browser can be
disabled via a write protection switch (DIP switch on the motherboard). When
hardware write protection is enabled, only read access to the parameters is possi-
ble.
Before servicing, shut off all power to the analyzer.
4900002277
5