Parameters That Can Always Be Modified Via The Local Display; Defining The Access Code Via The Web Browser - Endress+Hauser SpectraSensors J22 TDLAS Operating Instructions Manual

Gas analyzer
Hide thumbs Also See for SpectraSensors J22 TDLAS:
Table of Contents

Advertisement

J22 TDLAS Gas Analyzer
SpectraSensors, Inc.
If parameter write protection is activated via an access code, it can also only
be deactivated via this access code→  79.
The user role with which the user is currently logged on via the local display
is indicated by the Access status parameter. Navigation path: Operation →
Access status.

Parameters that can always be modified via the local display

Certain parameters that do not affect the measurement are excepted from
parameter write protection via the local display. Despite the user-specific access
code, they can always be modified, even if the other parameters are locked.

Defining the access code via the Web browser

1. Navigate to the Define access code parameter (→  83).
2. Max. Define a max. 4-digit numeric code as an access code.
3. Enter the access code again in the Confirm access code parameter (→  83)
to confirm the code.
 The Web browser switches to the login page.
If no action is performed for 10 minutes, the Web browser automatically
returns to the login page.
If parameter write protection is activated via an access code, it can also only be
deactivated via this access code →  39.
The user role with which the user is currently logged on via Web browser is
indicated by the Access status parameter. Navigation path: Operation → Access
status
Parameters for configuring
the local display
Format display
Constrast display
Display interval
4900002277
83