Protective Chassis And Ground Connections; Protective Ground Cable - Endress+Hauser SpectraSensors J22 TDLAS Operating Instructions Manual

Gas analyzer
Hide thumbs Also See for SpectraSensors J22 TDLAS:
Table of Contents

Advertisement

4900002277
Protective Chassis
and Ground Connec-
tions
14
1 Cover borehole for the securing screw
2 Securing screw to lock the cover
Before connecting any electrical signal or power, connect the protective and
chassis grounds (→  34).
Protective and chassis grounds must be of equal or greater size than any
other current-carrying conductors, including the heater located in the sample
conditioning system
Protective and chassis grounds to remain connected until all other wiring is
removed
Protective grounding wire current carrying capacity must be the same as the
main supply, at a minimum
Earth bonding/chassis ground shall be at least 10 AWG (6 mm

Protective ground cable

2
Analyzer: 2.1 mm
2
Enclosure: 6 mm
(10 AWG)
The grounding impedance must be less than 1Ω.
1
1
 1: Ground connections
1
Protective ground
(14 AWG)
1
J22 TDLAS Gas Analyzer
2
)
1
1
SpectraSensors, Inc.
A0029800

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SpectraSensors J22 TDLAS and is the answer not in the manual?

Questions and answers

Table of Contents