4900002277
Protective Chassis
and Ground Connec-
tions
14
1 Cover borehole for the securing screw
2 Securing screw to lock the cover
Before connecting any electrical signal or power, connect the protective and
chassis grounds (→ 34).
•
Protective and chassis grounds must be of equal or greater size than any
other current-carrying conductors, including the heater located in the sample
conditioning system
•
Protective and chassis grounds to remain connected until all other wiring is
removed
•
Protective grounding wire current carrying capacity must be the same as the
main supply, at a minimum
•
Earth bonding/chassis ground shall be at least 10 AWG (6 mm
Protective ground cable
2
•
Analyzer: 2.1 mm
2
•
Enclosure: 6 mm
(10 AWG)
The grounding impedance must be less than 1Ω.
1
1
1: Ground connections
1
Protective ground
(14 AWG)
1
J22 TDLAS Gas Analyzer
2
)
1
1
SpectraSensors, Inc.
A0029800