4900002277
Parameter
Date/time format
Parameter
Dew pt method 1
—
Dew pt method 2
—
Conversion type
Used if Dew pt is
enabled by select-
ing a method
above.
Pipeline press.
Used if Dew pt is
enabled by select-
ing a method
above.
56
Description
Sets the display unit for Date/time
format.
Setting the dew point
The dew point submenu configures parameters needed to perform a moisture dew
point calculation.
Navigation
"Setup" menu → Dew point
Dew point
Parameter overview with brief description
Prerequisite
Sets the method used to cal-
culate dew point tempera-
ture.
Sets the method used to cal-
culate dew point tempera-
ture.
Sets conversion type used to
calculate dew point tem-
perature.
Sets the pressure at which
the dew point temperature
is calculated.
User Entry
Unit Choose list
Dew pt method 1
Dew pt method 2
Conversion type
Pipeline press.
Description
•
•
•
•
•
•
•
•
•
Floating point
number
J22 TDLAS Gas Analyzer
Factory Setting
•
dd.mm.yy
hh:mm
•
mm/dd/yy
hh:mm am/pm
User Entry
Factory Setting
ASTM1
ASTM2
ASTM2
AB
ASTM1
Off
ASTM2
AB
Ideal
Ideal
Real
•
50000 mbar a
•
725 psi a