J22 TDLAS Gas Analyzer
Parameter
SCC
Methane CH4
Ethane C2H6
Propane C3H8
IButane C4H10
N-Butane C4H10
Isopentane C5H12
N-Pentane C5H12
Neopentane C5H12
Hexane+ C6H14+
Nitrogen N2
Carbon diox. CO2
Hydrog.sulf. H2S
SpectraSensors, Inc.
Parameter overview with brief description
Description
Enables or disables the Stream
Change Compensation feature.
Sets the mole fraction of Methane
in the dry gas mixture.
Sets the mole fraction of Ethane in
the dry gas mixture.
Sets the mole fraction of Propane
in the dry gas mixture.
Sets the mole fraction of Ibutane in
the dry gas mixture.
Sets the mole fraction of N-Butane
in the dry gas mixture.
Sets the mole fraction of Isopen-
tane in the dry gas mixture.
Sets the mole fraction of N-Pen-
tane in the dry gas mixture
Sets the mole fraction of Neopen-
tane in the dry gas mixture
Sets the mole fraction of Hexane+
in the dry gas mixture
Sets the mole fraction of Nitrogen
in the dry gas mixture.
Sets the mole fraction of Carbon
dioxide in the dry gas mixture.
Sets the mole fraction of Hydrogen
sulfide in the dry gas mixture.

Carrying out additional display configurations

In the Display submenu you can set all the parameters associated with the
configuration of the local display.
Navigation
"Setup" menu → Advanced setup → Display
Display
User Entry
On
Off
0.4 to 1.0 mol
0.0 to 0.2 mol
0.0 to 0.15 mol
0.0 to 0.1 mol
0.0 to 0.1 mol
0.0 to 0.1 mol
0.0 to 0.1 mol
0.0 to 0.1 mol
0.0 to 0.1 mol
0.0 to 0.55 mol
0.0 to 0.3 mol
0.0 to 0.05 mol
Format display
4900002277
Factory Setting
Off
0.75 mol
0.1 mol
0.05 mol
0 mol
0 mol
0 mol
0 mol
0 mol
0 mol
0 mol
0.1 mol
0 mol