Endress+Hauser SpectraSensors J22 TDLAS Operating Instructions Manual page 125

Gas analyzer
Hide thumbs Also See for SpectraSensors J22 TDLAS:
Table of Contents

Advertisement

J22 TDLAS Gas Analyzer
Diagnostic
Short Text
Number
110
Peak track adjust-
ment exceeded
111
Peak track adjust-
ment warning
Diagnostic of electronic
201
Device failure
242
Software incompatible
252
Modules incompatible
262
Sensor electronic
connection faulty
270
Main electronic failure Change main electronic module.
271
Main electronic failure
272
Main electronic failure
273
Main electronic failure Change electronic.
275
I/O module 1 to n
defective
276
I/O module 1 to n
faulty
283
Memory content
300
Sensor electronics
(ISEM) faulty
SpectraSensors, Inc.
Remedy Instructions
Check process.
1.
Check spectrum.
2.
Reset peak tracking.
3.
Check process.
1.
Check spectrum.
2.
Reset peak tracking.
3.
Restart device.
1.
Contact service.
2.
Check software.
1.
Flash or change main electronics
2.
module.
Check electronic modules.
1.
Change electronic modules.
2.
Check or replace connection cable
1.
between sensor electronic module
(ISEM) and main electronics.
Check or replace ISEM or main elec-
2.
tronics.
Restart device.
1.
Change main electronic module.
2.
Restart device.
1.
Contact service.
2.
Change I/O module.
Restart device.
1.
Change I/O module.
2.
Reset device.
1.
Contact service.
2.
Restart device.
1.
Replace sensor electronics.
2.
4900002277
Diagnostic
Status Signal
Behavior
(from factory)
(from factory)
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
119