4.9.2.3. TOTAL mode (total reset, start, stop)
Description
 The total is started, stopped or reset.
 Settable range: START, STOP, TOTAL RESET
START : Starts totalizing. Totalizes continuously from the stopped status.
STOP : Stops totalizing. Setting cannot be changed when it is not stopped.
RESET : Resets the total memory to the reset value, and starts totalizing.
<Note> A resetting action simultaneously resets both forward total memory and reverse total memory.
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.4.1.)
Operation
Reset the total value (reset value 0m
(example)
*Set the "TOTAL UNIT" beforehand. Refer to "4.9.4 Setting the LCD indication"
Key operation
Press the
Press the
ENT
Press the
Press the
ENT
Press the
ENT
Press the
Press the
ENT
Press the
ESC
the measurement mode.
3
), and restart a total.
Description
key twice to display "OUTPUT SETUP".
key once to display "ZERO ADJUSTMENT".
ENT
key for 5 times to display "TOTAL".
key once to display "TOTAL MODE".
ENT
key once to blink the cursor.
ENT
key twice to display "TOTAL RESET".
key twice to execute "TOTAL RESET".
ENT
――― The total operation is started. ―――
key twice and then press the
ESC
key for 3 times to enter
-40-
Display
0.00
m3/h
+
127.26
m3
OUTPUT SETUP
ZERO ADJUSTMENT
SET ZERO
TOTAL

TOTAL MODE

TOTAL MODE
TOTAL MODE
TOTAL RESET
TOTAL MODE
 COMPLETE 
TOTAL MODE
TOTAL RESET
0.00
m3/h
0.00
m3
START
START