4.9.1.3. Output limit
Description
Upper and lower limits can be set within the range of analog output 0.8mA
to 23.2mA (-20% to 120%).
Settable range
(1) Output lower limit: -20% to 0% (0.8mA to 4mA)
(2)
Output upper limit: 100% to 120% (20mA to 23.2mA)
For actual keying, refer to the typical operation indicated below. Set the
protection to OFF beforehand. (See Section 4.4.1.)
Operation
Set "-10% (2.4mA)" to lower limit, and "110% (21.6mA)" to upper limit.
(example)
* Set the piping parameters and "FLOW UNIT" beforehand.
Key operation
Press the
▼
Press the
ENT
▼
Press the
▼
Press the
ENT
▼
Press the
▼
Press the
ENT
▼
Press the
▼
Press the
▼
Press the
ENT
▼
▼
――― OUTPUT LIMIT LOW has been registered. ―――
▼
▼
Press the
▼
Press the
ENT
▼
Press the
▼
Press the
▼
Press the
ENT
▼
▼
――― OUTPUT LIM. HIGH has been registered. ―――
▼
▼
Press the
ESC
the measurement mode.
Description
key twice to display "OUTPUT SETUP".
key once to display "ZERO ADJUSTMENT".
ENT
key for 4 times to display "RANGE".
key once to display "KIND".
ENT
key for 7 times to display "OUTPUT LIMIT LOW".
key once to blink the cursor.
ENT
key once to align the cursor to "2".
key several times to set "1".
key once to register.
ENT
key once to display "OUTPUT LIM. HIGH".
key once to blink the cursor.
ENT
key once to align the cursor to "2".
key several times to set "1".
key once to register.
ENT
key twice and then press the
ESC
23.2mA
Lower limit
-20%
key for 3 times to enter
-36-
Analog output
20mA
4mA
0%
100% 120%
0.8mA
Display
OUTPUT SETUP
ZERO ADJUSTMENT
RANGE
KIND
FLOW RATE
OUTPUT LIMIT LOW
OUTPUT LIMIT LOW
OUTPUT LIMIT LOW
OUTPUT LIMIT LOW
OUTPUT LIMIT LOW
COMPLETE
↓
OUTPUT LIMIT LOW
OUTPUT LIM. HIGH
OUTPUT LIM. HIGH
OUTPUT LIM. HIGH
OUTPUT LIM. HIGH
OUTPUT LIM. HIGH
COMPLETE
↓
OUTPUT LIM. HIGH
0.000
m/s
0.000
m3/h
Upper limit
Flow rate
SET ZERO
-20 %
-20 %
-20 %
-10 %
-10 %
120 %
120 %
120 %
110 %
110 %