Calibration Settings; Calibration Selection - Panasonic HG-S Series User Manual

Contact-type digital displacement sensor
Hide thumbs Also See for HG-S Series:
Table of Contents

Advertisement

FUNCTION SETTINGS

5-7 Calibration Settings (

Structure of calibration setting menu
Long-press for 2 seconds

5-7-1 Calibration Selection (

You can perform zero-point adjustment and span adjustment when a sensor head is replaced to
reduce installation error.
Calibration settings
Calibration selection
(CAL.SEL)
Measure zero point.
Target point
Zero point
5-94
CALC: Only appears on master unit
Set values
Default (DEFAUL)
User settings (USER)
Acquired value 1 (CL.SET1)
Acquired value 2 (CL.SET2)
Adjustment value 2 (AJ.VAL2)
Step 1
)
[Calibration selection]
)
Set to the factory default state.
Execute calibration using the preset values be-
low, and correct the zero point and span.
Move the spindle to the zero point for calibration
and acquire the measured value.
Move the spindle to the target point for span ad-
justment and acquire the measured value.
Enter a number for the target value.
Step 2
Measure target point.
© Panasonic Industrial Devices SUNX Co., Ltd. 2019
Function
Step 3
Enter span target value.
Correct deviation between detected
value and target value.
Span