Replacing The Rubber Bellows (Regular Type Only) - Panasonic HG-S Series User Manual

Contact-type digital displacement sensor
Hide thumbs Also See for HG-S Series:
Table of Contents

Advertisement

APPENDIX

7-3 Replacing the Rubber Bellows (Regular Type Only)

● When replacing the rubber bellows, take care that no dirt or other contaminants get on
the spindle. Risk of malfunctioning. If any dirt gets on the spindle, wipe clean with abso-
lute alcohol. Do not allow the rubber bellows to become twisted during attachment.
● Note that the measuring force will vary depending on the attachment state of the rubber
bellows.
● The standard rubber bellows is a consumable part. Replace it regularly as a preventive
maintenance.The rubber bellows can deteriorate quickly depending on usage environ-
ment. If it deteriorates, it generates cracks and other problems, causing dust and water to
enter and resulting in a malfunction.
1.
Turn the probe screw in the direction of the arrow and remove the probe from the spindle.
When turning the probe screw, hold the cut face of the spindle with the provided sensor
head fastening wrench to prevent the spindle from turning.
Hold the sensor head fastening wrench in place, and turn only the probe.
Sensor head fastening
wrench (accessory)
2.
Remove the rubber bellows from the spindle.
3.
Fit the new rubber bellows onto the spindle.
Insert to the top edge of the spindle as shown below.
7-4
Rubber bellows (option)
Insert to top edge
© Panasonic Industrial Devices SUNX Co., Ltd. 2019