Parameters For The Plate Method In Setup; Immersing Speed (Mm/Min) - kruss K11 MK3 User Manual

Tensiometer
Table of Contents

Advertisement

Measurements with the plate or rod method
Rod
For the rod a smaller amount of test liquid is required due to the shorter wetted
length. However, the results are slightly less accurate when the wetted length is
shorter.
Determine the diameter of the rod as close to the lower edge as possible and with
an accuracy of at least 10 μm.
Multiply the measured diameter by the number π (3.1416) and enter the result as
the wetted length.
7.2

Parameters for the plate method in Setup

The first two pages of the
the measurement with the plate method. Assumingly, they rarely need to be altered.
Only those parameters which are relevant for the plate method are described here; the
Setup
complete
To open the
.
The parameters which are relevant for the plate method are
Search speed
setup menu.
7.2.1

Immersing speed (mm/min)

After a surface has been detected the plate is immersed in the liquid to a depth defined
by you. You should alter the immersion speed here when you are measuring very
viscous or viscoelastic substances. However, the value of 100 mm/min set by us is
normally suitable.
Setup
menu is described in chapter 10.
Setup
menu press the key
on the first page and
menu contain parameters which affect the course of
Immersion depth
page 41
Immersion speed
on the second page of the
and

Advertisement

Table of Contents
loading

Table of Contents