Operating the instrument
To switch the K11 to standby mode: Press the key
for at least two seconds.
4.1.3
Resetting the instrument
If the K11 no longer responds to the Control Panel: Press the
back of the K11.
4.1.4
Warming-up time
!
After the instrument has been switched on the force sensor must reach its
operating temperature before measurements can be started. This requires a
warming-up time of at least 2 hours.
In the standby mode the balance retains its operating temperature; this means that
no warming-up time is required when the instrument is switched on with the ESC
key. This is why the instrument should not be switched off completely during a
working day.
4.2
Error message behaviour
You will find a complete list of possible error messages in the appendix on page
106.
Switch off the K11 with the key
Remedy the error and switch on the K11 again.
.
page 21
RESET
button on the