Error messages

If an error occurs in controlling the instrument or when carrying out a measurement
then a corresponding error message will appear on the display. The possible error
messages and the steps to be taken to remedy them are given in the following list.
Error No.
17
18
19
33
34
Fault
Clock not
Occurs before the time has
working.
been set for the first time
and after the battery has
been exchanged.
Battery was
Only occurs immediately
empty and has
after the battery has been
been exchanged,
changed.
standard values
set.
Battery empty.
According to the
manufacturer the battery
has a life of up to 10 years;
however, usually the
battery has to be
exchanged earlier.
Force sensor not
Instrument is not steady
working (no data
enough.
transmission).
Overload (force
Appears if attaching and
acting
removing the weight when
downwards is too
adjusting the instrument.
large).
Reading limit
= 1
Reading limit
Appearance
Appendix
= 1
Remedy
Set the current time
and date (
ch. 10.4)
Only for information;
requires no action.
Does not appear the
next time the
instrument is switched
on.
Exchange the battery
in the K11 controller
PCB. Contact our
technical customer
service if necessary.
Find a vibration-free
location.
Remove load and
press ESC key.