kruss K11 MK3 User Manual page 48

Tensiometer
Table of Contents

Advertisement

page 46
Display
Searching
Immersing
Cover with the
light phase
Measuring
Displays the
individual measured
values
Finished
Displays the final
results:
Avarage (mN/m)
Standard
deviation (mN/m)
Temperature (°C)
Measurements with the plate or rod method
Action
automatic
automatic
Carefully overlay
the heavy phase
with the light phase,
best done by
running it down the
vessel side from a
pipette.
automatic
automatic
Explanation
Searches for the surface: the liquid
is moved upwards until the probe
"jump" is registered. The surface
position is stored.
Immerses the probe to the given
immersion depth.
Returns to the found surface
height.
Records measured values until the
required number of values has
been obtained.
The measurement is complete.
The data found are used to display
the mean value of the interfacial
tension, the standard deviation
and the temperature.