kruss K11 MK3 User Manual page 109

Tensiometer
Table of Contents

Advertisement

Appendix
Error No.
35
49
50
51
52
53
65
Fault
Underload (force
Appears if attaching and
pressing against
removing the weight when
balance from
adjusting the instrument or
below is too
if the motor presses the
large).
measuring body against the
balance.
On switch-on
Motor at lower
limit.
information for the operator.
During measurement
film tearing is recorded; the
sample carrier moves
downwards to the limit.
On switch-on
Motor at upper
limit.
information for the operator.
During measurement no
film tearing is recorded or
no surface is found. The
sample carrier moves
upwards to the limit.
Motor cannot be
A fault exists.
initialized.
Motor transmits
A fault exists.
no data.
Critical internal
Motor is overloaded; e.g.
motor fault
when an object gets
(overload)
jammed.
Lamella torn
Lamella tears when
maximum force is being
determined.
Appearance
: only
: no
: only
page 107
Remedy
Remove load and
press ESC key to
switch instrument off
and on.
Briefly
press the ESC
key.
The measurement is
terminated. Confirm
with ESC and restart
the measurement
Briefly
press the ESC
key.
The measurement is
terminated. Confirm
with the ESC key and
start the measure-
ment again. If
necessary, ensure
that the surface is
higher.
Contact our technical
support.
Contact our technical
support.
Remove the cause of
the overload.
Briefly
press the ESC
key and restart the
measurement. If the
message appears
again reduce the
measuring speed.