page 64
The following table shows the course of the measurement:
Display
Sample:
insert the sample
and enter a name
Only if defined in
Setup:
In succession
1
Field 2
,
,
User:
please enter a
name
Taring
Put the stick onto
the hook
Measuring
Density measurement
Action
Insert the sample
and use the keypad
to enter a sample
name.
Enter the required
contents in the text
fields
Field
Field 3
Use the keypad to
enter an operator
name.
automatic
Place the probe on
the holding device
using forceps.
automatic
Density measurement
Explanation
Maximum 22 characters
The text fields, whose number is
defined in Setup, are called up in
succession (max. 3) and are used
to describe the measurement. The
text fields can be named in
LabDesk and these fields assigned
to the corresponding data of the
current measurement.
Maximum 22 characters
Taring the holding device for the
probe
Determines the weight of the
probe in air